Adding Samples During a Run
AEROSET
®
Customer Training Guide
4-23
94858-103 — July 2003
Adding Samples on the Sample Carousel
When the Sample Carousel is Busy Sampling
When samples are being aspirated from the
Sample Carousel
, the
System status is RUN, and the Sample Carousel Movement Indicator
Light is illuminated or blinking.
1. Order the patient samples in the
Order Samples
screen.
2. Select
<PAUSE>
in the Action Area. The
RUN OPTIONS
screen
is displayed.
3. Select
<Pause>
in the right-hand column of the
RUN OPTIONS
screen to pause sampling.
WARNING:
If a diluted sample has been dispensed, the
Sample Arm may move after the status has changed to
PAUSE.
Wait 20 seconds before accessing the Sample
Carousel area
.
WARNING:
The AEROSET System does not stop the
movement of the Reagent 2 Probes, ICT™ Unit, Mixers,
or Cuvette Washers so the Assay Read Times are not
affected.
CAUTION:
The System completes the dispensing
and mixing of any sample and reagent already aspi-
rated. Confirm the PAUSE status in the System Sta-
tus Area and confirm that the R1 Indicator on the
LUI (Local User Interface) is illuminated before pro-
ceeding.
4. Wait for the System status to change to PAUSE and then place the
samples in the
Sample Carousel
in the specified positions.
NOTE:
For priority sampling, load the sample in the first
available position next to the position currently being
aspirated.
• To speed up when the sample is
aspirated, assign a C/P close to the
position currently being aspirated.
!
!
!
Summary of Contents for Aeroset
Page 1: ...Abbott Laboratories Abbott Park IL 60064 94858 103 July 2003 AEROSET Customer Training Guide...
Page 2: ...ii AEROSET Customer Training Guide 94858 103 July 2003 NOTES...
Page 14: ...xiv AEROSET Customer Training Guide 94858 103 July 2003...
Page 56: ...1 42 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 1 SYSTEM TOUR...
Page 90: ...2 34 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 2 BASIC RUN...
Page 152: ...MODULE 4 RUN OPTIONS 4 26 AEROSET Customer Training Guide 94858 103 July 2003...
Page 158: ...MODULE 4 RUN OPTIONS 4 32 AEROSET Customer Training Guide 94858 103 July 2003...
Page 201: ...Review Questions AEROSET Customer Training Guide 5 41 94858 103 July 2003 NOTES...
Page 204: ...MODULE 5 RESULT HANDLING 5 44 AEROSET Customer Training Guide 94858 103 July 2003...
Page 244: ...6 40 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 6 CALIBRATION...
Page 287: ...Review Questions AEROSET Customer Training Guide 7 43 94858 103 July 2003...
Page 290: ...7 46 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 7 QUALITY CONTROL...
Page 336: ...8 46 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 8 CONFIGURATION...
Page 337: ...AEROSET Customer Training Guide 9 1 94858 103 July 2003 MODULE 9 ICT UNIT MODULE 9 ICT UNIT...
Page 352: ...9 16 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 9 ICT UNIT...
Page 373: ...As Needed Maintenance AEROSET Customer Training Guide 10 21 94858 103 July 2003 NOTES...
Page 378: ...10 26 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 10 MAINTENANCE...
Page 415: ...AEROSET Customer Training Guide Glossary 1 94858 103 July 2003 GLOSSARY GLOSSARY GLOSSARY...
Page 450: ...Appendix 14 AEROSET Customer Training Guide 94858 103 July 2003 NOTES HAZARDS...
Page 451: ...AEROSET Customer Training Guide Index 1 94858 103 July 2003 INDEX...
Page 456: ...Index 6 AEROSET Customer Training Guide 94858 103 July 2003 NOTES Index...