System Configuration
AEROSET
®
Customer Training Guide
8-11
94858-103 — July 2003
SYSTEM CONFIGURATION Screen
The following System Configuration options are defined on the
SYSTEM CONFIGURATION
screen:
Option
Description
Auto Rerun
When selected, the System automatically reorders assays
for those samples that meet user specified criteria (Rerun
Rules), on an “assay by assay” basis. To activate the Auto
Rerun feature, this option must be selected and assay-
specific Rerun Rules defined on the
Assay Confguration
screen. The default is
OFF
.
NOTE:
Samples are not automatically returned
to the sample arm unless the Auto Return option
is selected on the
RUN OPTIONS
screen.
NOTE:
For information about configuration of
Rerun Rules, refer to the
Assay Configuration
section in this
Module.
Run Suspend
Time
Allows the Operator to configure how long the System
waits (in minutes) once a run is complete. While in the
SUSPEND
status, an abbreviated run initialization is
used (when the next run is initiated). The default setting
is 10 minutes.
Enable Cal Edits
Allows the Operator to edit read values
(
i.e
., absorbances) for calibrators in the
ASSAY
STATUS
screen. The default setting is
ON
.
NOTE:
If configured to
ON
, only a
Log On
level of “super” can edit calibrator values.
Smart Sampling
When selected, the System determines the assay sampling
order to maximize throughput, and minimize the number
of empty cuvettes. The default is
ON
.
For information about configuration of the
sampling sequence, refer to
System
Configuration
in
Section 2, Installation
Procedures and Special Requirements
of the
AEROSET System Operations Manual
.
Saving Mode
When selected, the System minimizes sample usage. The
over-aspiration volume is aspirated once for the same
sample, when aspirating more than one assay, rather than
each time an assay is aspirated. The default setting is
ON
.
NOTE:
Over-aspiration is an additional volume
of sample aspirated before the actual sample
volume required for analysis. This prevents the
sample from being diluted with System water in
the probe.
Auto Rerun:
• To get System to automatically
locate and aspirate reruns
without operator
intervention, auto return must
be ON.
Summary of Contents for Aeroset
Page 1: ...Abbott Laboratories Abbott Park IL 60064 94858 103 July 2003 AEROSET Customer Training Guide...
Page 2: ...ii AEROSET Customer Training Guide 94858 103 July 2003 NOTES...
Page 14: ...xiv AEROSET Customer Training Guide 94858 103 July 2003...
Page 56: ...1 42 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 1 SYSTEM TOUR...
Page 90: ...2 34 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 2 BASIC RUN...
Page 152: ...MODULE 4 RUN OPTIONS 4 26 AEROSET Customer Training Guide 94858 103 July 2003...
Page 158: ...MODULE 4 RUN OPTIONS 4 32 AEROSET Customer Training Guide 94858 103 July 2003...
Page 201: ...Review Questions AEROSET Customer Training Guide 5 41 94858 103 July 2003 NOTES...
Page 204: ...MODULE 5 RESULT HANDLING 5 44 AEROSET Customer Training Guide 94858 103 July 2003...
Page 244: ...6 40 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 6 CALIBRATION...
Page 287: ...Review Questions AEROSET Customer Training Guide 7 43 94858 103 July 2003...
Page 290: ...7 46 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 7 QUALITY CONTROL...
Page 336: ...8 46 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 8 CONFIGURATION...
Page 337: ...AEROSET Customer Training Guide 9 1 94858 103 July 2003 MODULE 9 ICT UNIT MODULE 9 ICT UNIT...
Page 352: ...9 16 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 9 ICT UNIT...
Page 373: ...As Needed Maintenance AEROSET Customer Training Guide 10 21 94858 103 July 2003 NOTES...
Page 378: ...10 26 AEROSET Customer Training Guide 94858 103 July 2003 NOTES MODULE 10 MAINTENANCE...
Page 415: ...AEROSET Customer Training Guide Glossary 1 94858 103 July 2003 GLOSSARY GLOSSARY GLOSSARY...
Page 450: ...Appendix 14 AEROSET Customer Training Guide 94858 103 July 2003 NOTES HAZARDS...
Page 451: ...AEROSET Customer Training Guide Index 1 94858 103 July 2003 INDEX...
Page 456: ...Index 6 AEROSET Customer Training Guide 94858 103 July 2003 NOTES Index...