background image

C

I/

A

H

M

55

0

-E

N R

ev

. D

 

0

7.

20

18


We reserve the right to make technical changes or modify the contents of this document 
without prior notice. With regard to purchase orders, the agreed particulars shall prevail. 
ABB does not accept any responsibility whatsoever for potential errors or possible lack of 
information in this document.
We reserve all rights in this document and in the subject matter and illustrations contained 
therein. Any reproduction, disclosure to third parties or utilization of its contents – in 
whole or in parts – is forbidden without prior written consent of ABB.  
© ABB 2018 

3KXA835550R4401


ABB Limited 

 

 

Measurement & Analytics 

 

Oldends Lane, Stonehouse  
Gloucestershire, 
GL10  3TA  
UK  

 

Tel:  +44 (0)1453 826661
Fax:  +44 (0)1453 829671 
Email: [email protected]
 

ABB Inc.  

 

Measurement & Analytics 

 

125 E County Line Road
Warminster, PA 18974
USA
Tel: +1 215 674 6000
Fax: +1 215 674 7183

abb.com/measurement

Solution preparation

Reagent Solution – 

5M (20% W/V) Sodium Hydroxide

The solution used to fill the reagent container is made up as shown 
below – consumption is approximately 250 ml in 2 to 4 weeks.

1. Weigh out 2.5 (±0.1) g EDTA and transfer to a 500 ml (16.9 fl 

oz) measuring flask (a little high purity water can be used to 
help this transfer).

2. In a separate vessel, weigh out 100 (±1) g sodium hydroxide, 

NaOH, pellets (analytical reagent grade) and dissolve in 
approximately 300 ml (10.0 fl oz) high purity water in a plastic 
container. Allow this solution to cool.

3. Transfer this solution to the measuring flask, shake well to 

dissolve the EDTA and top up to the mark with more high 
purity water.

Standard solution

Choose a convenient value for the hydrazine concentration of the
standard solution – typically 50 or 80 µg kg

-1

. Other concentrations

can be used if required.

Prepare a stock solution of 1000 mg l

-1

 hydrazine as follows:

1. Weigh out 4.058 (±0.001) g analytical reagent grade 

hydrazine sulphate (N

2

H

4

,H

2

SO

4

) and dissolve in 

approximately 800 ml (27.0 fl oz) high purity water.

2. Transfer to a 1 l (33.81 fl oz) volumetric flask and top up to  

the mark with more high purity water.

3. Dilute the stock solution to provide the required standard 

solution for the particular measuring range (usually 30 or 
80 µg kg

-1

).

Warning.

Sodium Hydroxide is extremely caustic and must be 

handled with great care. Wear gloves and eye protection.

Warning.

Hydrazine sulphate is an irritant to skin and eyes. 

Avoid breathing the dust. Wear gloves, eye protection and a 
dust mask when handling this substance.

Note.

Hydrazine solutions deteriorate with time: replace the 

stock solution at monthly intervals. Dilute standard solutions 
must be freshly prepared.

Summary of Contents for Navigator 550

Page 1: ...e the correct solutions prior to commissioning to purchase solutions contact your local ABB representative to prepare solutions refer to Solution preparation overleaf For more information Further publ...

Page 2: ...ndling containers reservoirs and delivery systems that contain chemical reagents and standards Protective eye wear and hand wear must always be used when contact with chemicals is possible Normal safe...

Page 3: ...posal systems In conformity with European local and national regulations European electrical equipment users must now return old or end of life equipment to the manufacturer for disposal at no charge...

Page 4: ...les within flexible earthed metal conduit 3 Mounting the wet section Refer to Fig 2 for wet section dimensions The wet section weighs 4 5 kg 10 lb excluding solutions Referring to Fig 2 1 Mark the wal...

Page 5: ...on site All other internal wet section tubing connections are factory made Referring to Fig 44 1 Fill the standard solution 2 high and reagent containers with the correct solutions refer to Solution p...

Page 6: ...L situated on the top of the reagent dosing chamber M 7 Place a small beaker into the main case and hold the end of reagent tube K over it to enable the reagent to flow through the tube to displace an...

Page 7: ...Press the key below the Select prompt on the Standard Solution page 18 Enter the required concentration of the standard solution using the keys to scroll to the required numbers Press the key below th...

Page 8: ...used to help this transfer 2 In a separate vessel weigh out 100 1 g sodium hydroxide NaOH pellets analytical reagent grade and dissolve in approximately 300 ml 10 0 fl oz high purity water in a plast...

Reviews: