background image

6

CI/AHM550–EN Rev. D

Setup

This section describes how to set up the analyzer up for first-time
use.

If multiple wet-sections are being connected to 1 transmitter, an
additional setup procedure is required – refer to Operating
Instructions (OI/AHM550-EN).

Transmitter start-up

Ensure all electrical connections have been made and switch the
power

on

to

the

transmitter.

If

the

wet-section

is

being

commissioned for the first time, calibration and programming of
parameters is required – complete steps 1 to 13, then follow the
instructions from step 14 onwards to complete setup at the
transmitter.

Setting up the wet-section

Referring to Fig. 5:

1. Fill the hydrazine sensor assembly with fresh gel as follows:

a. Holding the black closure cap tightly in place on the filling

syringe, snap the syringe plunger into position and remove
the black closure cap.

b. Slowly inject the filling gel through the lower hole in the

outer jacket until it reaches the top hole.

c. Remove the syringe and replace its closure cap.

d. Fit lower 10-32 UNF blanking plug

A

.

e. Fit upper 10-32 UNF blanking plug

B

.

2. Push the hydrazine sensor assembly

C

into clip

D

on the

sub-panel, ensuring sensor outlet tube

E

is positioned

directly above drain tundish

F

.

3. Connect the tube

G

from the mixing / delay coil to inlet

nipple

H

.

4. Connect the hydrazine sensor

I

(red) and temperature

sensor

J

(blue) electrical connectors to the appropriate

color-coded sockets on the hydrazine sensor connector
block.

5. Set up an appropriate sample flow into the wet-section –

check that the solution weirs over the adjustable overflow pipe
in the constant-head unit and flows through the sensor.

6. Remove reagent tube

K

from membrane clamp

L

, situated

on the top of the reagent dosing chamber

M

.

7. Place a small beaker into the main case and hold the end of

reagent tube

K

over it to enable the reagent to flow through

the tube to displace any air bubbles.

8. Reconnect reagent tube

K

to membrane clamp

L

.

9. Remove mixing / delay coil tubing

G

from reagent dosing

chamber outlet

N

.

10. Establish flow through the new disc by clamping the sample

inlet tube

O

and applying suction from a plastic syringe to

the reagent dosing chamber outlet

N

.

11. Refit the mixing / delay coil tubing

G

to the reagent dosing

chamber outlet

N

.

12. Remove the clamp from sample inlet tube

O

.

13. Allow approximately one hour for caustic dosing to be

established (the pH of the effluent at the sensor outlet must be
at least 10.5).

Analyzer accuracy is dependent on the flow rates of the
standard and sample through the sensor. The flow rates have
been pre-set at the factory but, to ensure accuracy, it is
advisable to check the flow rates and adjust if necessary –
refer to the Operating Instructions (OI/AHM550-EN) for flow
rate values.

At the transmitter:

14. Calibrate the transmitter as described in the Commissioning

instructions (CI/AWT540-EN).

15. Access the Configuration level menus as follows:

a. Press the

key (below the

icon).

The

Operator

menus are displayed:

Caution.

Do not attempt to set up the analyzer unless the

wet-section and transmitter are fully installed and ready for
operation.

Warning.

The wet-section uses chemical solutions in its

operation. Suitable precautions must be taken to ensure safe
handling.

Note.

Hold the sensor firmly at the top so that the centre

portion is not pushed out when the tube is connected.

CAL

2012–04–15

10 : 34: 52

4.28

4.28

°C

21.7

21.7

Navigator

ppb

N2H4

Tag 1

2012–04–15

10 : 34: 52

4.28

4.28

°C

21.7

21.7

Navigator

Operator Pages

Data Views
Logs
Alarm Acknowledge
Manual Hold
Autoscroll
Media Card
Enter Configuration

ppb

Summary of Contents for Navigator 550

Page 1: ...e the correct solutions prior to commissioning to purchase solutions contact your local ABB representative to prepare solutions refer to Solution preparation overleaf For more information Further publ...

Page 2: ...ndling containers reservoirs and delivery systems that contain chemical reagents and standards Protective eye wear and hand wear must always be used when contact with chemicals is possible Normal safe...

Page 3: ...posal systems In conformity with European local and national regulations European electrical equipment users must now return old or end of life equipment to the manufacturer for disposal at no charge...

Page 4: ...les within flexible earthed metal conduit 3 Mounting the wet section Refer to Fig 2 for wet section dimensions The wet section weighs 4 5 kg 10 lb excluding solutions Referring to Fig 2 1 Mark the wal...

Page 5: ...on site All other internal wet section tubing connections are factory made Referring to Fig 44 1 Fill the standard solution 2 high and reagent containers with the correct solutions refer to Solution p...

Page 6: ...L situated on the top of the reagent dosing chamber M 7 Place a small beaker into the main case and hold the end of reagent tube K over it to enable the reagent to flow through the tube to displace an...

Page 7: ...Press the key below the Select prompt on the Standard Solution page 18 Enter the required concentration of the standard solution using the keys to scroll to the required numbers Press the key below th...

Page 8: ...used to help this transfer 2 In a separate vessel weigh out 100 1 g sodium hydroxide NaOH pellets analytical reagent grade and dissolve in approximately 300 ml 10 0 fl oz high purity water in a plast...

Reviews: