background image

CI/AHM550–EN Rev. D

5

Connecting the sample lines

To make external sample inlet connections:

1. Fit a shut-off valve (not supplied) at each sample inlet.

Referring to Fig. 3:

2. Connect the sample inlet tubing

A

using flexible PVC tubing

(

1

/

4

in. ID non-flowmeter or

3

/

8

in. ID with flowmeter) to barbed

connector

B

at the base of:

sample inlet

C

(wet-section without flowmeter)

or

sample inlet

D

(wet-section with flowmeter)

3. Ensure the drain tubing

E

is kept as straight as possible and

is routed to a suitable drain.

4. Ensure the 2 drain tubes

F

from the constant-head unit

G

are kept as straight as possible and routed through the funnel

H

at the base of the wet-section. Drain tubes are

factory-fitted but may have moved in transit.

Connecting the standard solution

2 (high) and reagent tubing

Connections to the standard solution 2 (high) and reagent
containers (at the QD couplings) must be made on site. All other
internal wet-section tubing connections are factory-made.

Referring to Fig. 44:

1. Fill the standard solution 2 (high) and reagent containers with

the correct solutions – refer to Solution preparation.

2. Connect the standard solution 2 (high) tubing QD coupling

plug

A

to the mating connector

B

at the base of the

standard solution 2 (high) container.

3. Connect the reagent solution tubing QD coupling plug

C

to

the mating connector

D

at the base of the reagent solution

container.

Note.

Sample inlet tubing is customer-supplied.

Keep sample drains as short as possible and vertical to
enable the sample to drain freely.

Where particulate matter is present (for example,
magnetite in boiler samples) it is recommended that a 60
micron sample filter is fitted to the sample inlet line.

Fig. 3 Connecting the sample lines

Note.

If reagent and standard solutions have not already been

prepared, proceed to

Solution preparation

(overleaf).

Fig. 4 Connecting the standard solution 2 (high) and reagent tubing
QD couplings

Standard Solution 2

(High) tank – 400 ml

Reagent Solution

tank – 200 ml

Filler caps

Summary of Contents for Navigator 550

Page 1: ...e the correct solutions prior to commissioning to purchase solutions contact your local ABB representative to prepare solutions refer to Solution preparation overleaf For more information Further publ...

Page 2: ...ndling containers reservoirs and delivery systems that contain chemical reagents and standards Protective eye wear and hand wear must always be used when contact with chemicals is possible Normal safe...

Page 3: ...posal systems In conformity with European local and national regulations European electrical equipment users must now return old or end of life equipment to the manufacturer for disposal at no charge...

Page 4: ...les within flexible earthed metal conduit 3 Mounting the wet section Refer to Fig 2 for wet section dimensions The wet section weighs 4 5 kg 10 lb excluding solutions Referring to Fig 2 1 Mark the wal...

Page 5: ...on site All other internal wet section tubing connections are factory made Referring to Fig 44 1 Fill the standard solution 2 high and reagent containers with the correct solutions refer to Solution p...

Page 6: ...L situated on the top of the reagent dosing chamber M 7 Place a small beaker into the main case and hold the end of reagent tube K over it to enable the reagent to flow through the tube to displace an...

Page 7: ...Press the key below the Select prompt on the Standard Solution page 18 Enter the required concentration of the standard solution using the keys to scroll to the required numbers Press the key below th...

Page 8: ...used to help this transfer 2 In a separate vessel weigh out 100 1 g sodium hydroxide NaOH pellets analytical reagent grade and dissolve in approximately 300 ml 10 0 fl oz high purity water in a plast...

Reviews: