CI/AHM550–EN Rev. D
7
b. Press the
key to select the
Enter Configuration
menu
and press the
key (below the
icon).
The
Access Level
page is displayed:
c. Use the
key to scroll to the Advanced access level
and press the
key (below the
Select
prompt) to enter
the top level
Configuration
menus.
d. Use the
/
keys to scroll to the
Calibrate
page and
press the
key (below the
Select
prompt to enter the
Calibrate menus.
16. Use the
key to scroll to the
Calibration Setup
menu, press
the
key (below the
Select
prompt) to access the
Standard
Solution
menu.
17. Press the
key (below the
Select
prompt on the
Standard
Solution
page).
18. Enter the required concentration of the standard solution
using the
/
keys to scroll to the required numbers.
Press the
key (below the
Next
prompt) to save each
selected number. When the last number is highlighted, press
the
key (below the
OK
prompt) to set the value and exit
the page.
Fig. 5 Setting up the sample flow
D
E
C
Red hydrazine
sensor connector
(upper socket)
Blue hydrazine
temperature sensor
connector
(lower socket)
Blue
Red
Note.
If, during normal operation, the transmitter does
not display the expected hydrazine level, refer to the
Operating Instructions (OI/AHM550-EN) for
troubleshooting procedures.
Access Level
Back
Select
Read Only
Calibrate
Advanced
Service