background image

4

CI/AHM550–EN Rev. D

Sample requirements

The sampling point should be as near as possible to the wet-section
and must provide a thoroughly mixed representative sample.

The sample must conform to the following conditions:

Sample must contain less than 10 ppm suspended solids with
a particle size no greater than 60 µm.
(If particle sizes exceed 60 µm, use a 60 µm filter.)

Sample temperature must be within the range 5 to 55 °C
(41 to 131 °F).

Sample flow rates must be within the range 100 to
400 ml/min (6.10 to 24.4 to cu in./min.).

Sample must be at atmospheric pressure.

Locating the wet-section

Referring to Fig. 1:

1. Locate the wet-section / transmitter indoors in a clean, dry,

well-ventilated, easily-accessible area.

2. Do not locate in rooms containing corrosive gases or vapors –

for example, with chlorination equipment or chlorine gas
cylinders.

3. Select a location not subject to strong electrical and magnetic

fields. If this is not possible (for example, where mobile
communications equipment is present), use screened cables
within flexible, earthed, metal conduit.

Mounting the wet-section

Refer to Fig. 2 for wet-section dimensions. The wet-section weighs
4.5 kg (10 lb), excluding solutions.

Referring to Fig. 2:

1. Mark the wall using the dimensions shown.

2. Drill and plug 3 holes

A

and

B

in the wall suitable for M6

or

1

/

4

in. fixings.

3. Screw in top fixing

A

, leaving a gap of 20 mm (0.78 in.)

between the fixing head and the wall.

4. Hang

the

wet-section

onto

fixing

A

,

ensuring

the

wet-section is retained firmly against the wall.

5. Secure the wet-section to the wall using 2 fixings

B

.

Caution.

This equipment will be damaged if subjected to

freezing conditions.

Fig. 1 Locating the wet-section

Navigator 500

Navigator 500

Humidity

0 to 95 %

Avoid vibration

IP54

55 °C

(131 °F)

Max.

0 °C

(32 °F)

Min.

Maximum cable distance – transmitter to

wet-section 30 m (98 ft.)

Note.

Clearance – the enclosure doors can open 180°. If

mounting in a confined area, allow sufficient clearance for door
opening.

Note.

Fixing

A

cannot

be

adjusted

once

the

wet-section is placed over it.

Dimensions in mm (in.)

Fig. 2 Mounting the wet-section

 

Ø 7 (0.27)

210 (8.27)

480

(18.90)

185 (7.28)

290 (11.41)

2 Holes Ø 7

(0.27)

385

(15.16)

80.2

(3.15)

Allow sufficient clearance for

the door to open fully

281 (11.06)

Summary of Contents for Navigator 550

Page 1: ...e the correct solutions prior to commissioning to purchase solutions contact your local ABB representative to prepare solutions refer to Solution preparation overleaf For more information Further publ...

Page 2: ...ndling containers reservoirs and delivery systems that contain chemical reagents and standards Protective eye wear and hand wear must always be used when contact with chemicals is possible Normal safe...

Page 3: ...posal systems In conformity with European local and national regulations European electrical equipment users must now return old or end of life equipment to the manufacturer for disposal at no charge...

Page 4: ...les within flexible earthed metal conduit 3 Mounting the wet section Refer to Fig 2 for wet section dimensions The wet section weighs 4 5 kg 10 lb excluding solutions Referring to Fig 2 1 Mark the wal...

Page 5: ...on site All other internal wet section tubing connections are factory made Referring to Fig 44 1 Fill the standard solution 2 high and reagent containers with the correct solutions refer to Solution p...

Page 6: ...L situated on the top of the reagent dosing chamber M 7 Place a small beaker into the main case and hold the end of reagent tube K over it to enable the reagent to flow through the tube to displace an...

Page 7: ...Press the key below the Select prompt on the Standard Solution page 18 Enter the required concentration of the standard solution using the keys to scroll to the required numbers Press the key below th...

Page 8: ...used to help this transfer 2 In a separate vessel weigh out 100 1 g sodium hydroxide NaOH pellets analytical reagent grade and dissolve in approximately 300 ml 10 0 fl oz high purity water in a plast...

Reviews: