CAUTION!
For analyzer functionality and user safety, only use batteries
provided by ABB.
To replace the battery, perform the following (see Figure 67):
Figure 67: Battery Replacement
1. Ensure the analyzer power is OFF.
2. Open the case by unlatching two clamps next to the handle. (See Figure 4.)
3. Unplug the battery connector.
4. Holding the battery in place, unscrew the captive Phillips screw securing the
battery faceplate.
5. Rotate the faceplate downward.
6. Remove the battery from the case ensuring not to damage other
components.
7. Inspect the new battery for damage or bulging.
8. Place the new battery inside the faceplate and rotate the faceplate up.
9. Secure the faceplate to the battery with the captive screw.
10. Plug in the battery connector.
11. Close the case and fully lock the two clamps.
12. Power up the analyzer and verify proper installation.