ABB GLA131 Series Microportable Analyzers
User Manual
ABB PROPRIETARY INFORMATION
18
Gas Inlet and Outlet
Gas to be measured is connected to the ¼ inch push-connect
IN
port.
The acceptable inlet gas-pressure range is 0 to 15 psig.
The ¼ inch push-connect OUT port may:
Remain open, expelling non-hazardous gas into the air,
Be routed to a different location or system, or
Be recirculated through an enclosed gas-flux chamber.
Power
The analyzer is operational for up to 3 hours when using the internal battery only.
The battery may be recharged using the AC/DC power supply to the DC power
inlet. This power supply contains a 10 A 32 V DC automotive fuse inside the external
power-input harness.
There is also an optional battery charger (ACC-MICRO-BC) when extra batteries
need to be charged outside the analyzer.
WARNING!
Do not use a battery other than the battery provided by ABB.
Data Interface
USB ports – Used to transfer data to a USB memory device or connect a USB
keyboard and mouse.
Serial port (9 pin D-sub) – For real-time digital measurement output.
Video port (15 pin D-sub) – Connects an external monitor to the analyzer.
MIU port (9 pin D-sub) – For connecting to a Multiport Inlet Unit (optional).
Provided with cable for connection to an MIU data port.
Ethernet port – Connects the analyzer to a local area network (LAN) and
allows access to the data directory using an external computer.
Modbus TCP - For reading analyzer process variables and health status via a
Modbus master using the Modbus TCP version of the protocol.