Environmental Requirements
Table 4 defines the environment requirements of the analyzer during operational
and non-operational conditions. Operation outside of these requirements may
cause damage to the analyzer.
Table 4: Environment Requirements
Requirements
Specification
Operational Ambient
Temperature
5 °C to 45 °C (41 °F to 113 °F)
Non-Operational
Temperature
4 °C to 50 °C (40 °F to 122 °F)
Relative Humidity
Below the dew point for the area, avoiding
condensation on surfaces
WARNING!
To prevent damage to the analyzer, do not expose the analyzer to,
or use the analyzer in, rain or snow.
CAUTION!
ABB does not endorse operation of any analyzer (including
MicroGuard™) not certified for operation in hazardous locations
continuously in air that contains methane levels higher than the
lower flammability limit (LFL), where LFL for methane in air is 4.4%
or 44,000 ppm.
1 For maximum lithium-ion battery life and safety, the recommended storage temperature is between 4 °C to 25 °C