ABB GLA131 Series Microportable Analyzers
User Manual
ABB PROPRIETARY INFORMATION
75
6
Service
WARNING!
Use of controls or adjustments, or performance of procedures
other than those specified, may result in hazardous radiation
exposure.
The analyzer laser is not field serviceable. Should a laser fail in the field, the
analyzer must be shipped back to ABB for service repair.
The analyzer laser is equipped with a magnetic interlock switch. Whenever the case
cover is opened, the magnetic switch breaks contact from the magnetically
encoded counterpart, and power is interrupted from the Laser Controller PCB in
the analyzer, thus, disabling the laser.
Service to the analyzer itself must be performed only by certified service personnel
trained on servicing this instrument. User adjustments inside the analyzer are not
necessary or recommended by the manufacturer.
Basic Service Procedure
WARNING!
Only ABB certified maintenance personnel may unlock and open
the analyzer housing or perform internal maintenance.
Before servicing the analyzer, perform the following procedure:
1. Turn off the analyzer power via the software.
2. Unplug the power supply or the battery kit.
3. Wait 10 seconds for the power to dissipate.
4. Make sure the analyzer is on a stable surface and on the right side (main
label facing up). Unlock the padlock and open the analyzer cover.
5. Make sure all LEDs are off.
Spare Parts, Maintenance, and Accessories
Mirror Cleaning Kit
Part number: MTN-CLEAN-1V
Spare Pump and Filters Kit
Part number: SPK-131V2-LITE