32
FSM4000
ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
… 6 Electrical connections
Connecting the sensor
Connecting the signal and magnet coil cables
The flowmeter sensor is connected to the transmitter via the signal / magnet coil cables (part no. D173D025U01 / D173D147U01). The
coils of the flowmeter sensor are supplied with a field voltage by the transmitter over terminals M1/M3. Connect the cables to the
sensor in accordance with the following drawing, using a screwdriver with proper size and width.
Terminal box model SE21, SE21F
Terminal box model SE41F
1
red
2
blue
3
yellow
4
white
5
black
6
SE clamp
7
signal cable
8
ground connection
9
magnetic coil cable
1
red
2
blue
3
yellow
4
white
5
black
6
SE clamp
7
signal cable
8
ground connection
9
magnetic coil cable
Figure 30: Connection to sensor
Terminal designation
Connection
1 + 2
Wires for the measuring signal (red, blue)
1S, 2S
Shielding for signal wires
U+, U-
Power supply for preamplifier via signal cable shielding
16
Wire for reference signal (white)
3A
Shielding for reference signal cable
3
Measuring ground (yellow)
M1 + M3
Connections for magnetic field excitation (black)
SE
Outer cable shield