130
FSM4000
ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
… 13 Specification
… Sensor SE21 / SE21F
Storage temperature
−
25 to 70 °C (-13 to 158 °F)
IP rating in accordance with EN 60529
• IP 67
• IP 68 (Option)
Pipe vibration according to EN 60068-2-6
Transmitter
• In the range of 10 to 55 Hz, max. deflection 0.15 mm
Sensor
• In the range of 10 to 55 Hz, max. deflection 0.15 mm
• In the range of 10 to 55 Hz, max. acceleration 2 g
Material loads for process connections
Limits for allowable fluid temperature (TS) and allowable
pressure (PS) are a function of the liner and flange material used
(see the factory tag and name plate of the unit).
Devices with variable process connections / wafer type SE21
DN 3 to 100 (
¹⁄₁₀
to 4”)
Process connection
PFA liner
Nominal diameter
DN
PS
max
[bar]
TS
min
TS
max
Wafer type
3 to 50
(¹⁄₁₀
to 2”)
65 to 100
(2½ to 4”)
40
16
−
40 °C
(-40 °F)
130 °C
(266 °F)
Welded spuds
3 to 40
(¹⁄₁₀
to 1½”)
50; 80
(2”, 3”)
65, 100
(2½ to 4”)
40
16
10
−
25 °C
(−13 °F)
130 °C
(266 °F)
Threaded pipe
connection conforming
to DIN 11851
3 to 40
(¹⁄₁₀
to 1 ½”)
50; 80
(2”, 3”)
65, 100
(2½ to 4”)
40
16
10
−
25 °C
(−13 °F)
130 °C
(266 °F)
Tri-Clamp conforming to
DIN 32676
3 to 50
(¹⁄₁₀
to 2”)
65 to 100
(2 ½ to 4”)
16
10
−
25 °C
(−13 °F)
121 °C
(250 °F)
Tri-Clamp in acc. with
ASME BPE
3 to 100
(¹⁄₁₀
to 4”)
10
−
25 °C
(−13 °F)
130 °C
(266 °F)
External threads ISO 228 3 to 25
(¹⁄₁₀
to 1”
16
−
25 °C
(−13 °F)
130 °C
(266 °F)