Updating the Analyzer Software
The analyzer’s software must be updated twice a year. The software expires in June and
December. About six weeks before the current software expires, your facility will receive
a software update packet with complete instructions on how to perform the update.
15 days before the software expires, the analyzer will display “CLEW Expiring, Update
Required”. If you see this message and you have not received the update packet, call
Abaxis Technical Service at 1-800-822-2947.
Software updates require a VetScan i-STAT Serial Downloader or Serial Downloader/
Recharger and a PC with Windows 2000 or higher.
Thermal Probe Check
The analyzer’s thermal probes should be checked every six months. This check can be
performed in conjunction with the analyzer software updates.
Cleaning the Analyzer
If the analyzer is placed on a wet surface or if any liquid is spilled on it, dry
immediately. The analyzer may be damaged if liquid enters the battery compartment,
cartridge port or case.
Clean the display screen and case using a gauze pad moistened by any of the following:
•
A mild non-abrasive cleaner
•
Detergent
•
Soap and water
•
Alcohol
•
10% bleach solution
Avoid getting excess fluids in the seam between the display screen and the case. Rinse
using another gauze pad moistened with water and dry.