Check/Change Date and Time
1. Press to turn on analyzer
2. Press to change screen to Administration Menu
3. Press (Set Clock)
4. Press (No password is required)
5. Use or to move the cursor to a digit if it
needs to be changed. Use number keys to change digit.
Note: Analyzer uses a 24 hour clock.
Therefore, 1:00 PM is displayed as 13:00.
6. Press to accept changes, or to cancel changes.
MENU
MENU
5
ENT
3
ENT
3
4
4
13:26 18JUNyy
Administration Menu
1 - Analyzer Status
2 - Data Review
3 - Quality Tests
4 - Customization
5 - Set Clock
6 - Transmit Data
7 - Utility
Enter Current
Time And Date
13:36
06/18/13
mm/dd/yy
ENTER - Set And Exit
MENU - Cancel
Check Software and Status
1. From the Administration Menu, press
(Analyzer Status)
2. Check that software (CLEW & Version) in the
Analyzer is up to date.
1
15:26 18JUNyy
Administration Menu
1 - Analyzer Status
2 - Data Review
3 - Quality Tests
4 - Customization
5 - Set Clock
6 - Transmit Data
7 - Utility
Analyzer Status
Temp: 27.1˚C
Pressure: 761mmHg
Battery: 8.54V
Uses: 100
Serial: 30144-B
CLEW: A12
Version: JAMS121B
Custom: Default0
Stored Records
Total: 1
Unsent: 1
Temp: Analyzer’s reading of room temperature.
Analyzer will function between 16-30˚C.
Pressure: Analyzer’s reading of barometric pressure.
Analyzer will function between 300-1000mmHg.
Battery: Analyzer’s reading of battery voltage.
Uses: Total number of cartridge runs on analyzer.
Serial: Serial number of analyzer.
CLEW: Standardization software installed on analyzer.
Maintains accuracy of tests.
Version: Application software installed on analyzer.
Directs functions.
Custom: Name of customization profile in analyzer.
Default0 indicates default settings. Analyzer can be
customized for certain site-specific and cartridge
testing characteristics.
Art: 723085-00K Rev. Date: 30-OCT-2014
7