HW V4 / © 2005 – 2013 Aaronia AG, DE-54597 Strickscheid
32
Manual Spectran V4
Functional description
5.3.19. BackBB
(change bargraph display mode, option 20x)
Use BackBB to change the display mode of the bargraph
(only if option 20x installed)
Selecting “Spec” will display the highest value of the
spectrum display (default) Selecting “BBand” will display the
Peak-Power of the power meter (option 20x).
The power meter allows extremely fast broadband detection
of RF signal sources in dBm or dBµV, with frequencies of up
to 10GHz (12GHz). Detection sensitivity in this mode is
around -50dBm (preamp on), depending on frequency. The
maximum allowable signal level is approximately +10dBm.
Further information can be found in chapter 8.10.
5.3.20. AntTyp
(configure connected antenna)
The AntTyp setting must be configured correctly to reflect
the type of antenna connected, as otherwise, mis-
measurements can occur. For example, the “HL7025”
setting corresponds to the HyperLOG7025 and HyperLOG
4025, “HL7040” to the HyperLOG7040 and so forth.
„HL7025“ = HyperLOG7025
„HL7040“ = HyperLOG7040
In case no antenna is connected, please use the “None”
(0dBi) option.
You can also create your own custom antenna types using
our MCS software.