background image

 

10

 

Measurements 

Measurement With The Desired Systolic Pressure 

  If your systolic pressure is expected to 

exceed 230 mmHg, use this procedure. 

1. Place the cuff on the arm   

(preferably the left arm).   

2.  Press and hold the 

 START 

 button to 

display the last data number. 

3.  Press and hold the 

 START 

 button 

again until a number about 30 to 40 

mmHg higher than your expected 

systolic pressure appears.   

4. Release the 

 START 

 button to start 

measurement, when the desired number 

is reached. Then continue to measure 

your blood pressure as described on the 

previous page. 

 
 
 
 

Notes For Proper Measurement 

  Sit down in a comfortable position. Place the arm to be used for the measurement 

on a table or other support so that the center of the cuff will be at the same height 

as your heart. 

  Relax before taking a measurement. If you are excited or depressed by emotional 

stress, the measurement will reflect this stress as a higher (or lower) than normal 

blood pressure reading and the pulse reading will usually be faster than normal. 

  An individual's blood pressure varies constantly, depending on what you are doing 

and what you have eaten. What you drink can have a very strong and rapid effect 

on your blood pressure. 

  This device bases its measurements on the heartbeat. If you have a very weak or 

irregular heartbeat, the device may have difficulty determining your blood pressure. 

  Should the device detect a condition that is abnormal, it will stop the measurement 

and display an error symbol. Refer to page 4 for the description of the symbols. 

  This device is intended for use by adults only. Consult with your physician before 

using this device on a child. A child should not use this device unattended. 

 

Refer to the previous page for 

measurement 

At heart level

Inflation in progress

Release the button when 

stopping inflation and 

starting measurement. 

Press and hold

Last data number

Press and hold

Summary of Contents for UA-621

Page 1: ...Digital Blood Pressure Monitor Model UA 621 Instruction Manual 1WMPD4001898D...

Page 2: ...ring Data 11 What Is An Irregular Heartbeat 12 Pressure Bar Indicator 12 WHO Classification Indicator 12 About Blood Pressure 13 What Is Blood Pressure 13 What Is Hypertension And How Is It Controlled...

Page 3: ...es X ray or other devices with strong electrical fields Used equipment parts and batteries are not treated as ordinary household waste and must be disposed of according to the applicable local regulat...

Page 4: ...pressure oxygen chamber or an oxygen tent It may cause a fire or explosion Parts Identification Part of Display Pressure Bar Indicator and WHO Classification Indicator MEMORY Battery Indicator I H B...

Page 5: ...ry The battery power indicator during measurement Low Battery The battery is low when it blinks Replace all batteries with new ones when the indicator blinks Unstable blood pressure due to movement du...

Page 6: ...ory automatically 2 Recalling Data Refer to page 11 When nothing is displayed press and hold the START button to display the data number The data number of the most recent measurement and the data are...

Page 7: ...ambient temperature and may be shorter at low temperatures Use the specified batteries only The batteries provided with the device are for testing monitor performance and may have a limited life Remo...

Page 8: ...cm Adult cuff CUF D A Arm size The circumference at the biceps Note Model UA 621 is not designed for using a small adult cuff Symbols printed on the cuff Symbols Function Meaning Recommended Action A...

Page 9: ...ent Place the center of the cuff at the same height as your heart Remain still and keep quiet during measurement Do not measure immediately after physical exercise or a bath Rest before taking the mea...

Page 10: ...p inflation at any time press the START button again 3 When inflation is complete deflation starts automatically and the heart mark blinks indicating that the measurement is in progress Once the pulse...

Page 11: ...tional stress the measurement will reflect this stress as a higher or lower than normal blood pressure reading and the pulse reading will usually be faster than normal An individual s blood pressure v...

Page 12: ...1 When nothing is displayed press and hold the START button to display the data number The data number of the most recent measurement and the data are displayed 2 The data are automatically displayed...

Page 13: ...Note We recommend contacting your physician if you see this indicator frequently Pressure Bar Indicator The indicator monitors the progress of pressure during measurement WHO Classification Indicator...

Page 14: ...uce salt and fat intake Have regular physical checkups Maintain proper weight Why Measure Blood Pressure At Home Blood pressure measured at a clinic or doctor s office may cause apprehension and can p...

Page 15: ...batteries are drained completely the mark does not appear Replace all batteries with new ones The cuff is not fastened properly Fasten the cuff correctly You moved your arm or body during the measure...

Page 16: ...Type UA 621 Measurement method Oscillometric measurement Measurement range Pressure 20 280 mmHg Pulse 40 180 beats minute Measurement accuracy Pressure 3 mmHg Pulse 5 Number of measurements Approximat...

Page 17: ...33C Input 100 240V Output 6V 500mA TB 233BF Input 240V Output 6V 500mA Accessories sold separately Arm size The circumference at the biceps AC adapter Catalog Number Plug Outlet type TB 233C Type C TB...

Page 18: ...ingdon Business Park Abingdon Oxfordshire OX14 1DY United Kingdom Telephone 44 1235 550420 Fax 44 1235 550485 A D Australasia Pty Ltd 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 6...

Reviews: