IMMUNITY TEST LEVELS: Enclosure Port
Phenomenon
Immunity test levels
Electrostatic discharge
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated RF EM fields
IEC 61000-4-3
10 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3
See table (Test specifications for
ENCLOSURE PORT IMMUNITY
to RF wireless communications
equipment)
Rated power frequency magnetic
fields
IEC 61000-4-8
30 A/m
50 Hz/60 Hz
IMMUNITY TEST LEVELS: PATIENT COUPLING Port
Phenomenon
Immunity test levels
Electrostatic discharge
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
IMMUNITY TEST LEVELS: Signal input/output Port
Phenomenon
Immunity test levels
Electrostatic discharge
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transients/bursts
IEC 61000-4-4
±1 kV
100 kHz repetition frequency
Conducted disturbances induced
by RF fields
IEC 61000-4-8
3 V 0.15 MHz - 80 MHz
6 V in ISM and amateur radio bands
between 0.15 MHz and 80 MHz
80% AM at 1 kHz