65
For the A-BPM and ANBP and ASBP modes of S-BPM, only the
current blood pressure measurement can be suspended and the
"
1 scope
" is performed at the next
start time
. Set the
AUTO switch to "
OFF
" to suspend A-BPM.
Remove the cuff if the current blood pressure measurement
cannot be stopped using the START/STOP switch.
Press the START/STOP switch to stop the measurement.
The "
1 scope
" is still performed at the next
start time
for
A-BPM and the ANBP and ASBP modes of S-BPM.
When pain in the arm or an unexpected condition occur,
stop the measurement, remove the cuff and consult a doctor.
Set the AUTO switch to "
OFF
" to suspend A-BPM.
Set the AUTO switch to "
ON
" to resume A-BPM automated
measurement. The
mark is shown on the LCD and OLED.
The recording of data continues until switched to "
OFF
".
How to use manual measurement during A-BPM
The procedure for temporary measurement that is not included in
preset program.
Step 1. If the OLED display is hidden, press the
START/STOP or EVENT switch to return to the A-BPM
waiting mode display.
Step 2. Press the START/STOP switch to immediately measure
the blood pressure during A-BPM.
Step 3. Measurement results are stored in the memory.
When the START/STOP switch is pressed during
measurement, the measurement is suspended.
Caution