59
8.4.1.
S-BPM Items and Parameters
Program
Item
Parameter
S-BPM program
Program
OBP , AOBP, HBP, ANBP, ASBP
Office blood pressure
OBP
N.A.
N.A.
Automated office blood pressure
AOBP
Measurement count
2 ,
1 to 5 counts
Interval time
5 ,
3 to 10 minutes
Home blood pressure
HBP
Measurement count
2 ,
1 to 5 counts
Interval time
1 ,
1 to 5 minutes
Automated night blood pressure
ANBP
Start time
2 ,
0 to 23 hours
#1
Measurement count
2 ,
1 to 5 counts
Interval time
1 ,
1 to 5 minutes
Automated self-blood pressure
ASBP
Start time of alarm
7 , 22 , 0 to 23 hours
#2
Measurement count
2 ,
1 to 5 counts
Interval time
1 ,
1 to 5 minutes
Enclosed characters
:
Factory settings.
#1
:
24 hours can be separated to six sections in maximum.
Each section can specify the
start time
of the blood pressure
measurement.
#2
:
24 hours can be separated to six sections in maximum.
Each section can specify the
start time of alarm
for the
blood pressure measurement.