8.2.3.
Initial Pressurization Value of S-BPM
An initial pressurization value of S-BPM can be set beforehand.
160, 180, 210, 240, 270, AUTO [mmHg]
If AUTO is specified, a proper initial pressurization value is
selected automatically. The factory setting is 180 mmHg.
Step 1. Set the AUTO switch to "
OFF
".
Step 2. If the display is hidden, press the START/STOP
or EVENT switch to return to the waiting mode display.
Step 3. Select a pressurization value using the START/STOP switch.
8.3.
A-BPM Preset Programs
The initial settings can be configured using the following methods.
Using the switches on the recorder.
Using a
dedicated peripheral
that is connected to the recorder
using the USB cable.
A-BPM can only be used while the automated measurement can be
performed.
Note
Specify the
start time
and
interval
calculated from the time
that the AUTO switch is set to "
ON
".
It is necessary to specify them again when using another A-BPM.
AUTO switch to "
OFF
"
S-BPM program
Pressurization value
Current time