32
6.2.1.
S-BPM Programs
Office blood pressure
OBP
When the / switch is pressed, blood pressure is measured
once and is stored in the memory.
Automated office blood pressure
AOBP
When the / switch is pressed, AOBP is started.
First, the device is idle for the interval time to allow the patient to relax.
Next, the AOBP performs a "
1 scope
".
"
1 scope
" consists of a set of blood pressure measurements and
interval times, which are repeated for the number of times of the
measurement count. The last interval time is omitted.
When the / switch is pressed during the "
1 scope
", the
"
1 scope
" is stopped.
T
:
Interval time
Press the / switch to start the measurement.
Time
Blood pressure measurement
Press the / switch to start the "
1 scope
".
Press the / switch to stop the "
1 scope
".
T
T
Time
T
T
T
Not measured
Press the / switch to start the "
1 scope
".
T
T
Time
T
T
T
1 scope
Blood pressure measurement