84
9.3.
Periodic Inspection
Perform the daily periodic inspection to use the recorder correctly.
The inspection is described below:
9.3.1.
Battery Pre-installation Inspection
Items
Description
Exterior
No damage or deformation from dropping.
No dirt, rust and scratches on any part.
No cracking or rattling of the panel.
Operation
No damage or rattling of switches and buttons.
Display
No dirt or scratches on the display panel.
Measurement
Cuff
The air hose must not be folded. If air remains in
the cuff, it may cause peripheral dysfunction from
stopping the blood flow in the arm.
The cuff bladder is correctly inserted inside the cuff
cloth.
No fraying of the cuff. The cuff doesn't ravel.
Replace the cuff when a problem is found.
The cuff is disposable.
If there is a crack or adhesive matter in the
connection between the cuff and cuff bladder.
If the air hose loses its flexibility and becomes
hard.
When the surface of the air hose becomes glossy
or feels oily.
When the air bladder has cracks.
We recommend replacing cuffs every three years,
regardless of frequency of use.
Wearing tools No damage in the carry holder, belt and cuff.
Connection
The air plug is connected to the air socket correctly.