71
Display Error code
Description
EC, E03
Timeout error
If the timeout parameter is set to “
t-Up1
”, the balance
did not receive the next character of a command within
the time limit of one second.
Confirm the communication.
EC, E04
Excess characters error
The balance received excessive characters in a
command.
Confirm the command.
EC, E06
Format error
A command includes incorrect data.
e.g. The data is numerically incorrect.
Confirm the command.
EC, E07
Parameter setting error
The received data exceeds the range that the balance
can accept.
Confirm the parameter range of the command.
Other errors
If the errors described above can not be released or
other errors are displayed, contact the local A&D dealer.
20-3 Asking For Repair
If the balance needs service or repair, contact the local A&D dealer.
The balance is a precision instrument. Use much care when handling the balance and observe
the following when transporting the balance.
Use the original packing material.
Remove the weighing pan from the main unit
Summary of Contents for GF-1000
Page 7: ...5 ...