50
Recalling the memory data
Confirm that the “Data memory (
data
)” parameter is set to “
2
”.
1 Press and hold the PRINT key until
reCall
is displayed.
2 Press the PRINT key to enter the memory recall mode. Recall the data in memory using the
following keys
RE-ZERO key
To proceed to the next data set.
MODE key
To go back to the previous data set.
PRINT key
To transmit the current data using the RS-232C interface.
With SAMPLE held down, press the CAL key
To delete the data currently displayed.
Note: Deleting the data will not increase the number of data that can be stored.
CAL key
To exit the memory recall mode.
3 Press the CAL key to return to the weighing mode.
Transmitting all memory data at one time
Confirm that the “Serial interface (
5if
)” parameters are set properly. For details, refer to “10.
FUNCTION TABLE” and “17. CONNECTION TO PERIPHERAL EQUIPMENT”.
1 Press and hold the PRINT key until
reCall
is displayed.
2 Press the SAMPLE key to display
out
.
3 Press the PRINT key to display
Sout no
with “
no
” blinking.
5 Press the RE-ZERO key to display
out go
with “
go
” blinking.
6 Press the PRINT key to transmit all data using the RS-232C interface.
7 The balance displays
Clear
when all data is transmitted. Press the CAL key to return to the
weighing mode.
Deleting all memory data at one time
1 Press and hold the PRINT key until
reCall
is displayed.
2 Press the SAMPLE key several times to display
Clear
.
3 Press the PRINT key to display
SClr no
with “
no
” blinking.
4 Press the RE-ZERO key to display
SClr go
with “
go
” blinking.
5 Press the PRINT key to delete all data
6 The balance displays
end
, then
reCall
when all data is deleted.
7 Press the CAL key to return to the weighing mode.
Summary of Contents for GF-1000
Page 7: ...5 ...
Page 79: ...77 23 EXTERNAL DIMENSIONS ...