37
Interval memory mode
The weighing data is periodically stored in memory.
Required setting
dout prt
3
Interval memory mode
dout data
2
Data memory function is used.
Stores weighing data.
dout int
Interval
time
Example
“For periodical weighing without a computer command and to output
all of the data, to a computer, at one time”
10-6 Description of the Item “Data format”
A&D standard format
5if type 0
This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B is
used, set the printer to MODE 1 or 2.
z
This format consists of fifteen characters excluding the terminator.
z
A header of two characters indicates the balance condition.
z
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus
sign is used.
z
The unit, consisting of three characters, follows the data.
DP (Dump print) format
5if type 1
This format is used when the peripheral equipment can not receive the A&D format. If an AD-8121B
is used, set the printer to MODE 3.
z
This format consists of sixteen characters excluding the terminator.
z
A header of two characters indicates the balance condition. No overload header is used.
z
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is
not zero or overloaded.
z
The unit, consisting of three characters, follows the data.
Unit
Data
Header
S T , + 0 0 0 1 2 . 7 8
1 1
g
CR LF
Terminator