Best operating temperature is between 20°C~25°C / 68°F~77°F at about 50%~60%
relative humidity. There shouldn’t be large temperature fluctuations.
The weighing room should be kept clean and dry.
The weighing table must be of a solid construction.
Corners of rooms are best as they are less prone to vibrations.
Don’t install the scale near heaters or air conditioners.
Don’t install the scale in direct sunshine.
Try to ensure a stable AC power supply when using the AC adapter.
Keep equipment containing magnets away from the scale.
Warm up the scale before use or leave it on standby overnight.
Ground the scale chassis for electrostatic discharge if the weighing conditions warrant.
Calibration
Calibration of the scale is required when the scale is initially installed, or if a remote
scale is added. Please refer to “
9
. CALIBRATION
” for more calibration information.
6