10-2.
F-Functions
“
” designates factory settings.
F-00-X Weighing Unit
Weight Display.
0
kg (kilograms, FC-
i
series), g (grams, FC-S
i
series)
1
3
lb (pounds).
Unit Weight (when “lb” is selected).
0
lb as piece weight.
1
3
lb as 1,000 piece weight.
F-01-X Operations
Operation Mode.
0
3
Normal operation. All features and keys available.
1
Simplified operation. The unit weight registration is by
sample only. All other keys are disabled.
“Add” Sample Request Override.
If the sample weight is too light and the scale asks to “Add”
more sample pieces, using this F-Function, the unit weight
can be entered without adding the requested sample pieces.
Or disable the “Add” sample request function.
0
“Add” sample request function is disabled. Light unit weight
can be accepted without “Add” more sample request.
1
3
The unit weight can be entered without requested “Add”
sample pieces (via the
ENTER
key).
2
The unit weight cannot be entered without requested “Add”
sample pieces (via the
ENTER
key).
Auto
REMOTE SCALE
After Unit Weight Registration.
The scale can be set to the remote scale after the unit weight
registration, instead of having to manually press the
REMOTE SCALE
key. Please note that this does not affect
any other use of the
REMOTE SCALE
key.
0
3
No automatic switching.
1
Automatic switching to the remote scale.
2
Automatic switching to the main scale.
f-01-01
f-01-02
f-01-03
F-01-03 requires
OP-05 and a
remote scale
f-00-01
USA Version ONLY
f-00-02
36