Entering the density of a liquid
Two ways to set the density of a liquid are available in the function table, "Liquid density input (
ldin
): by
entering the water temperature or by entering the density directly.
1 Press
the
MODE
key as necessary to select the density mode.
Press and hold the
MODE
key (for approx. 2 seconds) to select to "water
temperature" or "liquid density". In the density mode : The unit is "g (gram)". The
processing indicator (
)
blinks. Response adjustment is off.
If you cancel the density mode, press the
MODE
key.
Entering the water temperature
(
ldin
0
)
2 Press and hold the
MODE
key (for approx. 2 seconds). The water
temperature currently set (unit: °C, factory setting : 25°C) is
displayed. Use the following keys to change the value.
RE-ZERO
(+)key ....To increase the temperature by one degree.
(0°C is displayed after 99°C)
MODE
(-)key..........To decrease the temperature by one degree.
(99°C is displayed after 0°C)
key............To store the new water temperature, display
end
and return to the density mode. Proceed
to Step 1.
CAL
key...............To cancel the change and return to the density mode.
Proceed to Step 1.
The relation between the water temperature and density
°C
+0 +1 +2 +3 +4 +5 +6 +7 +8 +9
0
0.99984 0.99990 0.99994 0.99996 0.99997 0.99996 0.99994 0.99990 0.99985 0.99978
10
0.99970 0.99961 0.99949 0.99938 0.99924 0.99910 0.99894 0.99877 0.99860 0.99841
20
0.99820 0.99799 0.99777 0.99754 0.99730 0.99704 0.99678 0.99651 0.99623 0.99594
30
0.99565 0.99534 0.99503 0.99470 0.99437 0.99403 0.99368 0.99333 0.99297 0.99259
40
0.99222 0.99183 0.99144 0.99104 0.99063 0.99021 0.98979 0.98936 0.98893 0.98849
g/cm
3
Entering the density directly
(
ldin
1
)
Press and hold the
MODE
key (for approx. 2 seconds). The density
currently set (unit : g / cm
3
, factory setting : 1.0000g / cm
3
) is
displayed.
Use the following keys to change the value.
The range to set the density is 0.0000g / cm
3
to 1.9999g / cm
3
.
RE-ZERO
(+)key.....To increase the value of the digit selected.
MODE
(-)key ..........To decrease the value of the digit selected.
RANGE
key..........To select the digit to change the value.
key............To store the change, display
end
and return to the density mode. (Step 1).
CAL
key...............To cancel the change and return to the density mode. (Step 1)
Press and hold
(approx. 2 seconds)
Press and hold
(approx. 2 seconds)