BM series
35
7. Response
Adjustment
This function stabilizes the weight value by reducing the influence on weighing that is caused
by drafts and/or vibration at the place where the balance is installed. The adjustment of the
function can be performed by analyzing the environment automatically or hand-operation.
The state of the function has three stages as follows :
Indicator Parameter Response
Stability
FAST
Cond
0
MID.
Cond
1
SLOW
Cond
2
Fast response, Sensitive value
Slow response, Stable value
7.1.
Automatic Response Adjustment
This function automatically updates the response adjustment by analyzing the influence of
the environment using the internal mass.
1 Press and hold the
MODE
key (for approx. 2 seconds)
until
RESPONSE
is displayed, and then release the key.
2 The balance automatically sets the response characteristic.
Caution Do not allow vibration or drafts to affect the
balance during adjustment.
3 After automatic adjustment, the balance displays
end
,
returns to the weighing mode and displays the updated
response indicator for about thirty seconds.
Notes
If the automatic response adjustment fails, the balance
displays
CH
ng
. Check the ambient conditions such
as breeze and vibration, also check the weighing pan.
Then, perform adjustment again. Press the
CAL
key
to return to the weighing mode.
If there is anything on the weighing pan, the balance
displays
CH
e
. Remove the substance from the pan.
Press the
CAL
key to return to the weighing mode.
If the automatic response adjustment is not helpful, try
"
7.2. Manual Response Adjustment
".
Press and hold
(approx. 2 seconds)
Release
Result
Response indicator
Summary of Contents for BM-5