64
BM
series
Calibration test report using an external weight
Note
Calibration test does not perform adjustment.
If the data memory function is used (except
data
0
), the "Title block" and "End block"
cannot be output.
Key operation
1 Press and hold the
CAL
key to display
CC
out
and release the key.
2
Cal
0
is displayed.
3 To update the target weight value, press the
RANGE
key and proceed to Step 4.
To use the preset target weight value, proceed to Step 5.
4 Specify the target weight value using the following keys.
RANGE
key ..........To select the blinking digit.
RE-ZERO
(
+
)key ....To increase the value of the blinking digit.
MODE
(
-
)key.........To decrease the value of the blinking digit.
key............To store the new value.
5 Press
the
key. The zero point is measured and the weight value is displayed for a few seconds.
6 Place the displayed target weight on the pan. Press the
key to measure the weight. The weight
value is displayed for a few seconds.
7
end
is displayed. Remove the weight from the pan.
8 If GLP output is used,
glp
is displayed and the calibration test report is output.
9 The balance returns to the weighing mode.
Function
setting
"
info
1
" Function
setting
"
info
2
"
AD-8121
format
General format using the balance built-in clock data
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR
Carriage
return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration test type
Zero point value
Target weight value
Target weight
Signature
-----------
A
-
&
-
D
MODEL
-----
BM-300
S/N
-----
01234567
ID
------
LAB-0123
DATE
--
2011/02/20
TIME
----
12:34:56
CAL.TEST(EXT.)
ACTUAL
-------
0.0000
--
g
----
+200.0002
--
g
TARGET
----
+200.0000
--
g
SIGNATURE
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-