BM series
39
8.3.
Calibration Using the Internal Mass
This function calibrates the balance using the internal mass.
The only operation required is to press the
CAL
key.
1 Connect the AC adapter and warm up the balance for at
least one hour with nothing on the weighing pan.
2 Press
the
CAL
key to display
Cal
in
.
3 The balance performs calibration using the internal mass.
Do not allow vibration or drafts to affect the balance.
4
If
GLP output is set,
glp
is displayed, the calibration report is
output to the RS-232C interface and is stored in memory. Refer to
"GLP output (
info
)" and "Data memory (
data
)" of the function
table in "
11.2. GLP Report
" and "
12. Data Memory
".
end
is displayed after calibration.
5 The balance will automatically return to the weighing mode after calibration.
6 Confirm
weighing
accuracy using calibration test (
CC
in
).
GLP output