50
BM
series
Class Item
and
Parameter
Description
off
No limit
t-Up
Timeout
on
1 second
Selects the wait time to receive
a command.
off
No output
5if
Serial
interface
erCd
AK, Error code
on
Output
AK: ASCII code 06h
0
Water temperature
d5
fnc
Density
function
ldin
Liquid density input
1
Liquid density
Available only when density
mode is selected
Unit
Unit
Refer to "
5. Weighing Units
".
C5
in
Internal mass correction
Displayed only when the internal mass value correction
switch is set to
1
. Refer to "
8. Calibration
".
id
ID number setting
Refer to "
11. ID Number and GLP Report
".
0
3 seconds
1
10 seconds
ion
Neutralization time
2
Manual operation (max. 10 min.)
Refer to "
6.5. Ionizer
".
off
The
ION
key is not used.
8922
AD8922A control
on
Assigns the
ION
key to the
MODE
key on AD-8922A.
0
Assigns the
key and
RE-ZERO
key.
1
Assigns the
key and
ION
key.
ionfnc
Ionizer
function
e-5
External control input
2
Assigns the
ION
key and
RE-ZERO
key.
off
Not used
door
Door sensor
function
d-dt
Door sensor
on
Used
Refer to "
6.1. Basic Operation
"
0
Not used
1
Used (Limits weighing operations)
locfnc
Password
function
pa55
#4
Password function
2
Used (Allows basic weighing)
Refer to "
15. Password
Function”
.
ADM
IN
Password input by the administrator
loc
no.
Password
U5e
R
01
U5e
R
10
Password input by user 1
Password input by user 10 #5
: Factory settings.
#4 Only “Not used (
0
)” and “Used (
1
)” are available for balances with software version 1.50 to 1.64.
All the parameters are available for balances with software version 1.660 or later.
#5 Available for balances with software version 1.660 or later.
For balances with software version 1.50 to 1.64, only the password input by the administrator is
available.
10.3.
Description of the Class "Environment, Display"
Condition (
Cond
)
Cond
0
This parameter is for sensitive response to the fluctuation of a mass value. Used for
powder target weighing, weighing a very light sample or when quick response weighing
is required. After setting, the balance displays
FAST
.
Cond
2
This parameter is for stable weighing with slow response. Used to prevent a weight
value from drifting due to vibration or drafts. After setting, the balance displays
SLOW
.
Note In automatic response adjustment, this parameter is selected automatically.