10.2.
Details of the Function Table
Class Item
and
Parameter
Description
0
Fast response, sensitive value
FAST
1
MID.
Cond
Condition
2
Slow response, stable value
SLOW
Can be changed by
response adjustment.
0
Stable range is ±1 digit
1
±2
digit
#1
5t-b
Stability band width
2
Stable range is ±3 digits
The stabilization indicator
illuminates when the display
fluctuation is within the range.
off
OFF
ploc
Display lock function
on
ON
Display lock function to measure the
volume dispensed from the pipette.
0
OFF
1
Normal
2
Strong
trc
Zero tracking
3
Very strong
Keeps zero display by tracking
zero drift.
0
5 times/second
5pd
Display refresh rate
1
10 times/second
Period to refresh the display
0
Point (.)
pnt
Decimal point
1
Comma (,)
Decimal point format
off
OFF
p-on
Auto display-ON
on
ON
Turns on the weighing mode display
when the AC adapter is connected.
off
OFF
ba5fnc
Environment
Display
beep
Beep
on
ON
Sounds a beep when a key is
touched.
Cl
adj
Clock
Refer to "
10.7. Clock and Calendar Function
"
The time and date are added to
the output data.
0
Key mode
Accepts the
key only
when the display is stable.
1
Auto print mode A
Reference = zero
2
Auto print mode B
Reference = last stable value
Outputs data when the display is
stable and conditions of
ap-p
,
ap-b
and the reference value are
met.
prt
Data output mode
3
Stream mode /
Interval mode
With
data
0
, outputs data
continuously; with
data
1
, uses
interval memory.
0
Plus only
Displayed value > Reference
1
Minus only
Displayed value < Reference
ap-p
Auto print polarity
2
Both
Regardless of displayed value
0
10 digits
1
100 digits
ap-b
Auto print
difference
2
1000 digits
Difference between reference
value and displayed value
#2
0
Not used
1
Stores the weighing data
dout
Data output
data
Data memory
2
Stores the calibration data
Related items:
prt,
int
,
d-no
,
5-td
,
info
: Factory settings.
#1 The unit of readability is digit.
Example: If 1 mg display is selected using the
RANGE
key for the
BM-300
, 1 mg is one digit.
#2 Usable readability of the balance is one digit.
Example: In gram display, one digit is 0.01 mg for the