13.4. CF-function list
CF- Item
Setting value
Default User
setting
000 Decimal point position
0:
1:
2:
3:
0
0.0
0.00
0.000
(No decimal point)
1
001 Unit
0:
1:
kg
t
0
002 Push zero range
0:
1:
±2% of the maximum capacity
±10% of the maximum capacity
0
003 Reserved
1
004 Zero
tracking
0:
1:
2:
Gross is 0 when displaying gross
Gross is 0
Gross is 0 or display is 0
2
005 Power-on
zero
0:
1:
Disable
Enable
1
006 Execution Preset tare
0:
1:
Enable
Disable
1
007 When outputting the net value, the preset
tare value is output as well.
0:
1:
Disable
Enable
1
008 Output on over load and unstable state.
0:
1:
Disable
Enable
0
009 Reserved
2
010 Unit number of digits and decimal point for
output data
0:
1:
2:
3:
Two digits, "." dot
Three digits, "." dot
Two digits, "," comma
Three digits
,
"," comma
0
011 Accumulation
function
0:
1:
Disable
Enable
0
012 Overload condition on negative gross value. 0:
1:
Gross < - Maximum capacity
Gross <
-
20d
1
013 Manually print at negative gross value.
0:
1:
Enable
Disable
1