6.8. Multi-interval function
The multi-interval function divides weighing range into two or three ranges and can change scale interval
automatically according to the load applied.
Set the following items to use multi-interval function.
・
1st range maximum capacity and interval
・
2nd range maximum capacity and interval
・
3rd range maximum capacity and interval (if necessary)
The setting value must
meet the following conditions:
・
1st range capacity and interval < 2nd range capacity and interval < 3rd range capacity and interval
・
Example1.
1st range:
Capacity 20.00kg, Interval 0.01kg
2nd range:
Capacity 50.00kg, Interval 0.02kg
Setting value
3rd range:
Capacity 100.0kg, Interval 0.1kg
When the weighing value is between 0 and 20kg, 1st range
(
Interval 0.01kg
)
is selected.
When the weighing value is between 20 and 50kg, 2nd range
(
Interval 0.02kg
)
is selected.
Display
When the weighing value is between 50 and 100kg,3rd range
(
Interval 0.1kg
)
is selected
Example 2. The following is net display with Example 1 settings, when the tare value is 40kg.
When the net value is between -40 and -20kg, 2nd range (Interval 0.02kg) is selected.
When the net value is between -20 and 20kg, 1st range (Interval 0.01kg) is selected.
When the net value is between 20 and 50kg, 2nd range (Interval 0.02kg) is selected.
Display
When the net value is between 50 and 60kg, 3rd range (Interval 0.1kg) is selected.
1st range
+
2nd range
+
3rd range
< 10000
Capacity
Interval
Capacity
Interval
Capacity
Interval
Gross
Interval 0.1kg
Interval 0.02kg
Interval 0.01kg
Net
0kg 20kg 40kg
100kg
60kg
Interval 0.02kg
Tare
0kg 20kg
50kg
-40kg 60kg
-20kg
Interval 0.1kg
Interval 0.02kg
Interval 0.01kg
0kg 20kg
50kg
100kg
Gross