3.2. Functions
■
Digital zero (Zero adjustment)
Sets the measured value to zero using the ZERO key, external input or the command
signal.
The zero value is saved in non-volatile memory (EEPROM).
Available setting range
The available setting range is 1 to 100% of the rated capacity.
■
Zero tracking function
Automatically updates the zero point by sensing the zero point drift.
Available setting range
Tracking time :0.0 to 5.0 (second)
Tracking width :0.0 to 9.9 (d)
■
Power on zero function
Sets the zero state at power on.
■
Digital filter function
Cutoff frequency range: 0.7 to 11 (Hz)
■
Comparator function
HI, OK or LO is determined by setting the upper or lower limit value.
■
Hold function
Select from sample hold, peak hold, bottom hold, or bipolar peak hold.
■
Latch function
Perform latching of the displayed value, comparator output, analog output, serial
output, using the external LATCH input.
■
Relay output function (option)
Provides contact output of the HI, OK or LO signals.
■
Serial input and output function (option)
The RS-232C/RS-485 option can output measured values and input commands.
■
Analog output function (option)
Outputs the measured value as an analog voltage or current.