![A&D AD-4531B Instruction Manual Download Page 56](http://html1.mh-extra.com/html/aandd/ad-4531b/ad-4531b_instruction-manual_3079333056.webp)
56
16.5. Serial communication
Function No.
Setting range
Description
Default value User setting
f3-01
2400
to 38400
Baud rate:
2400
: 2400 bps
19200
: 19200 bps
4800
: 4800 bps
38400
: 38400 bps
9600
: 9600 bps
2400
f3-02
7 to 8
Data bit length:
7
:7 bits
8
:8 bits
7
f3-03
0 to 2
Parity:
0
: None
1
: Odd
2
: Even
2
f3-04
1 to 2
Stop bit:
1
: 1 bit
2
:2 bits
1
f3-05
1 to 2
Terminator:
1
: CRLF ASCII code: CR: 0D, LF: 0A
2
: CR
1
f3-06
1 to 6
Output mode:
1
: Stream
2
: Manual print mode
3
: Auto print mode
(Outputs data once when the measured
value exceeds the zero range and is
stabilized for the first time.)
4
: Auto print mode
(Outputs data each time the measured
value exceeds the zero range and is
stabilized.)
5
: Command mode
6
: Jet stream mode
(Outputs data at each sampling,
depending on the baud rate.)
2
f3-07
00 to 99
Instrument No.
When
00
, the ID is not added.
00
f3-08
6 to 8
Number of characters in measurement:
6
: 6 charactor
7
: 7 charactor
8
: 8 charactor
Including decimal point and polarity.
8
Summary of Contents for AD-4531B
Page 59: ...Blank...