16. SETTING LIST
When performing maintenance, use the following list as a memo.
When making inquiries about the product, inform your local A&D dealer of the user settings.
16.1. Calibration (C function)
Function No.
Setting range
Description
Default value User setting
Cf-01
0 to 5
Decimal point position of the measured value:
0
:
000000 3
:
000.000
1
:
00000.0 4
:
00.0000
2
:
0000.00 5
:
0.00000
0
Cf-02
1 to 50
Minimum division(d) of the measured value.
1
:1
5
:5
20
:20
2
:2
10
:10
50
:50
1
Cf-03
1 to 999999
Rated capacity
Measurement is possible up to the value of this
setting plus 8 d (8 minimum divisions).
Decimal point position depends on
Cf-01
.
70000
Cf-04
-7.00000 to
7.00000
Input voltage from the load cell at zero point.
□.□□□□□
mV/V
0.00000
Cf-05
0.00001 to
9.99999
Input voltage from the load cell at span.
□.□□□□□
mV/V
3.20000
Cf-06
-999999 to
999999
Displayed value for span (measurement point -
zero point).
Decimal point position depends on
Cf-01
.
32000
Cf-07
0 to 100
Zero adjustment range.
Expressed as a percentage of the rated capacity
with the calibration zero point as the center.
100
Cf-08
0.0 to 5.0
Zero tracking time. (Unit: second)
When
0.0
, zero tracking is not performed.
0.0
Cf-09
0.0 to 9.9
Zero tracking width. (Unit: d)
When
0.0
, zero tracking is not performed.
0.0
Cf-10
0 to 2
Power on zero:
0
: Digital zero function off
1
: Perform digital zero
2
: Use state when the power is disconnected.
0
Cf-11
1 to 3
Zero operation:
1
: On with
2
: On with
+
3
: On by pressing and holding for more
than 1 second
1
Cf-12
1 to 2
Zero of the I/O input:
1
: On / Off depends on the I/O input
2
: Only digital zero on (no off)
1
* In all settings:
When it is on,
off with
+