14. SERIAL INPUT AND OUTPUT
14.1. Data output format
Instrument number
It is added when
f3-07
is set to anything other than
00
.
Header
In the case of measurement values within the displayable range,
when "WT," is over (display blank), "OL," is added as a header.
Measured value
Specify the number of characters (including the sign and
decimal point) of the measured value to be output in the
f3-08
setting within the range of 6 to 8 characters. Even if
the measured value is within the displayable range, if the
measured value output to the serial output exceeds the number
of characters specified by
f3-08
, the output will be the same
as when over.
Unit
The unit character set by
f4-00
is added. If
f4-00
is set
to
0
, the characters set in
f4-01
to
f4-05
are added.
Terminator
Terminator for output data. Select either
CR
LF
or
CR
with
f3-05
.
Data output format example
When the measured value is
within the displayable range
(with decimal point)
WT,±0123.45kg
CR LF
When the measured value is
within the displayable range
(no decimal point)
WT,±0012345kg
CR LF
When the measured value is
over (display blank)
OL,±9999.99kg
CR LF
When instrument number is
added
@NNWT,±0123.45kg
CR LF
N is the number set in
f3-07
When the specified character
is added to the unit
WT,±0123.45UUUUU
CR LF
U is the character code
set in
f4-01
to
f4-05
When
f3-08
is set to
7
WT,±123.45kg
CR LF
When
f3-08
is set to
7
(Over)
OL,±999.99kg
CR LF
The same applies if the
measured value
exceeds 7 characters.
When
f3-08
is set to
6
WT,±123.4kg
CR LF
When
f3-08
is set to
6
(Over)
OL,±999.9kg
CR LF
The same applies if the
measured value
exceeds 6 characters.
@ N N W T , ± 0 1 2 3 . 4 5 k g
CR
LF
Instrument
number
Header
Measured value
Unit Terminator