Type
Advantages
Disadvantages
Description
6-wire connection
(recommended)
The error is small
even when the load
cell cable is
extended, a thin load
cell cable is used, or
multiple load cells
are used.
Complicated wiring
Use a 6-wire
shielded cable when
a summing box is
used.
4-wire connection Simple wiring
The influence of the
load cell cable
resistance worsens
the temperature
coefficient.
Prone to the
influence of the
contact resistance of
the connector
The error increases
when the load cell
cable is extended or
multiple load cells
are used.
Precautions on performing the 4-wire connection
Be sure to connect terminals 7 and 6 (EXC+ and SEN+) and terminals 5 and 4
(SEN- and EXC-).
If the cable needs extending, use a cable with as large a cross-sectional area as
possible or keep the cable as short as possible.