14.2.6. CAL span command
Update the input voltage (
Cf-05
) at the span of the CAL using the data that is input
when the command is received.
Command
CS
Command example
CS
CRLF
Response example
CS
CRLF
14.2.7. Function reading command
Confirm the function setting.
Query “?” function number: four characters
Command example
?F123
CRLF
Response example
F123,±456789
CRLF
14.2.8. Function setting command
Change the function setting.
Function number: four characters, the polarity, set value: characters.
Command example
F123,±456789
CRLF
Response example
F123,±456789
CRLF