![A&D AD-4531B Instruction Manual Download Page 31](http://html1.mh-extra.com/html/aandd/ad-4531b/ad-4531b_instruction-manual_3079333031.webp)
31
8.3.6. Unit
Function No.
Setting range Function
Description
Default value
Setting type
f4-00
0 to 4
Unit
0
: Specify the unit character
(
f4-01
to
f4-05
)
1
: kg
3
: t
2
: g
4
: Ib
1
P
f4-01
f4-02
f4-03
f4-04
f4-05
00 to 7F
Unit character1
Unit character2
Unit character3
Unit character4
Unit character5
Unit character added to the serial output.
Set using the hexadecimal ASCII code.
00
:
none
00
(Hexadecimal)
D
8.3.7. Errors
Display
Cause
Remedy
ad?e
The data can not be acquired from
A/D converter.
Repair is required.
eepe
Correct data cannot be read from
EEPROM.
Perform initialization.
If the initialization does not clear the
error, repair is required.
Cale
Calibration data error.
Perform calibration.
dt?e
A set value is out of range.
Check set values and correct if
necessary.
8.3.8. ASCII code (20h
~
7Fh)
Hexadecimal
00 01 02 03 04 05 06 07 08 09 0A 0B 0C 0D 0E 0F
20
SP !
" # $ % & '
(
)
*
+
,
-
.
/
30
0 1 2 3 4 5 6 7 8 9
:
;
< = >
?
40
@ A B C D E F G H I
J K L M N O
50
P Q R S T U V W X Y Z
[
\
]
^
_
60
`
a b c d e
f g h
i
j
k
l m n
o
70
p q r s
t u v w x y z
{
|
}
~ DL
20
SP is space
7F
DL is DEL