8
COMMAND
NAME
DESCRIPTION
N
o
.
COMMAND SYNT
AX
Command
Argument 1
Argument 2
Argument 3
Argument 4
Response
Pulse Output
Sets and controls action of the
programmable Pulse Output cir
cuitr
y.
NOTE:
Unit/Pulse value has to be entered
in currently selected EU.
EU has to be not time based
It is recommended to set the unit/pulse
value equal to the meter maximum
fl ow in the same units per second
equivalent. This will limit the pulse to
no faster than one pulse ever
y second.
Example:
Maximum fl
ow rate:
600 liter/min (600 liter/min = 10 liters
per second)
If Unit/Pulse is set to 10 liters per
pulse, the output will pulse once ever
y
second (F=1 Hz).
Pulse active time in ms has to be at
least twice less than pulse period (1/F).
In this example any value between 50
DQG
500 ms will be accep
WDEOH
.
17
P
U
Set Units Per Pulse
Parameter
.
<V
alue>
(Unit/Pulse)
In current E.U.
(example:
10 litr/pulse)
PU:<value>
Example:
PU:10
T
Set Pulse active T
ime
in ms
<value>
[
25-3276
ms]
PT
:<value>
Example: PT
:100
D
(disable pulse
output)*
P:D
E
(enable pulse output)
P:E
Q
(read current pulse
output Queue value)
PQ:<value>
(number of pulses in Queue)
F
Set Flow Start value
<value>
(0.0–100.0%FS)
PF:1.0
S
(read setting status)
PS:Mode,FlowStart,
Unit/Pulse,PulseT
imeInter
val
Example:
PS:E,1.0,1.666,100