COMMAND
NAME
DESCRIPTION
N
o
.
COMMAND SYNT
AX
Command
Argument 1
Argument 2
Argument 3
Argument 4
Response
Flow Alarms
Sets / reads the parameters of the mass
fl ow alarms.
Note: High alarm value has to be more
than Low alarm value.
Meter Flow Alarm conditions:
Flow
t
High Limit
= H
Flow
d
Low Limit = L
Low < Flow < High
= N
Alarm Settings Reply parameters:
M
– mode (E/D)
Hv – High settings value
Lv
– Low settings value
A – Action Delay (sec)
L
– Latch mode (0-1)
P – Power Up delay (sec)
11
FA
C
(set Mass Flow
Alarm High
and Low limits
parameters)
<V
alue>
(high limit,
%F
.S.)
[0.1 – 110.0]
%F
.S.
<V
alue>
(low limit,
%F
.S.)
[0.0 – 109.9]
%F
.S.
FAC:40.10,20.50
A
(Flow Alarm
action delay in
sec.)
<V
alue>
[0-3600]
FAA:<V
alue>
Example: F
AA:5
E
(enable fl
ow
alarm)
FA:E
D
(disable fl
ow
alarm)*
FA:D
R
(
5
ead current
status)
FAR:N (no alarm)
FAR:H (high alarm)
FAR:L (low alarm)
S
(Read current
settings)
FAS:M,Hv
,Lv
,A,L,P
Example:
FAS:E,40.00,20.00,2,0,8
P
(Flow Alarm
Power Up delay
in sec.)
<V
alue>
[0-3600]
FAP:<V
alue>
Example: F
AP:60
L
(Latch mode)
<V
alue>
(0-disabled*)
(1-enabled)
FAL:<V
alue>
where:
V
alue = 0 – 1
Example: F
AL:0