7
COMMAND
NAME
DESCRIPTION
N
o
.
COMMAND SYNT
AX
Command
Argument 1
Argument 2
Argument 3
Argument 4
Response
Diagnostic
Events
Registers
Read/Set/Reset current value of the Diagnostic
Events Registers
Status Register:
(Read/Reset)
Mask Register:
(R/W)
Latch Register:
(R/W)
See list of the Diagnostic Events below:
0 CPU_TEMP_HIGH
0x0001
1 DP EE INIT ERROR
0x0002
2 AP EE INIT ERROR
0x0004
3 VREF_OUT_OF_RANGE
0x0008
4 FLOW ABOVE LIMIT
0x0010
5 AP OUT OF RANGE
0x0020
6 G TEMP OUT OF RANGE 0x0040
7 ANALOG OUT ALARM
0x0080
8 SER COMM F
AILURE 0x0100
9 MB COMM F
AILURE
0x0200
A EEPROM F
AILURE
0x0400
B AUTOZERO F
AILURE
0x0800
C AP T
ARE F
AILURE
0x1000
D DP PRESSURE INV
ALID
0x2000
E AP PRESSURE INV
ALID
0x4000
F F
A
TAL_ERROR 0x8000
6
D
E
NO ARGUMENT
(read Diagnostic
Events status
register)
DE:<V
alue>
Example: DE:0x0
R
(reset Diagnostic
Events status regis-
ter to 0x0000)
DER:0x0
M
(Read/Set Diag-
nostic Events Mask
register)
NO ARGUMENT
(read current
settings)
DEM:0x1
<V
alue>
0x0000-0x0FFF
Set new value
NOTE: all 6
characters are
required
DEM:0x101
L
(Read/Set Diag-
nostic Events Latch
register)
NO ARGUMENT
(read current
settings)
DEL:0x1
<V
alue>
0x0000-0x0FFF
Set new value
NOTE: all 6
characters are
required
DEL:0x101