Rating 

 

Model 

Specifications 

A B C D 

 

Output voltage 

2H 0~30V 

2H 0~30V 

2H 0~30V 

2H 0~30V 

Output current 

2H 0~3A 

2H 0~5A 

2H 0~3A 

2H 0~5A 

Output fixed 

5V,3A 

5V,3A 

5V,3A 

5V,3A 

Output on/off 

V/A initializes 

X X 

 

 

 

Dimensions: 245(W)H140(H)H345(D)mm 

Weight: Approx.8kg 

2-2 Operation Mode 

(1). Independent: Two independent outputs and 5V fixed output. 

Output from 0 to rating volts and 0 to rating amperes. 

  (2). Series:          Output from 0 to ± rating volts at rating amperes each. 

Output from 0 to double rating volts at rating amperes. 

(3). Parallel:      Output from 0 to double rating amperes at rating volts. 

2-3 Constant Voltage Operation 

(1). Output current range                      0~rating voltage continuously adjustable. 

(2). Regulation                                        Line regulation

0.01%+3mV. 

Load  regulation

0.01%+3mV (rating current

3A) 

Load  regulation

0.02%+5mV (rating current>3A) 

(3). Recovery time                             

100us (50% Load change, Minimum load 0.5A) 

(4). Ripple & Noise                           

1mV rms, 5Hz~1MHz. 

(5). Temperature coefficient      

300PPM/ C 

2-4 Constant Current Operation 

(1). Output current range           0~rating current continuously adjustable. 

(2). Regulation                                                        Line regulation

0.2%+3mA. 

Load regulation

0.2%+3mA. 

(3). Ripple current                                           

3mA rms. 

2-5 Tracking Operation 

(1). Parallel operation 

Regulation                            Line    regulation

0.01%+3mV. 

Load  regulation

0.01%+3mV (rating current

3A) 

0.02%+5mV (rating current>3A) 

(2). Series operation 

Regulation                                                    Line    regulation

0.01%+5mV. 

Load  regulation

300mV 

A. Positive and Negative supply (Fig.4-4). 

 

 

(2)  Series Tracking Operation 

Summary of Contents for Precision AC- DC Power Supply

Page 1: ...Dual Channel 0 30V 5A max or 0 60V single channel AA Portable Power Corp http www batteryspace com Address 860 S 19th St Unit A Richmond CA 94804 Tel 510 525 2328 Fax 510 439 2808 Email Sales battery...

Page 2: ...output range The front panel current controls can be used to establish the output current limit overload or shout circuit when the supply is used as a constant voltage source independent or tracking m...

Page 3: ...utput on off key to enter voltage and current setting C Adjust the output voltage to the desired level using the MASTER VOLTAGE controls D Turn off the power supply and the equipment to be powered dur...

Page 4: ...h its Kindred has not This function is that when the output has been cut off the both intending voltage and current can be preset though in model A B condition only the intending voltage can be adjust...

Page 5: ...peration 1 Output current range 0 rating voltage continuously adjustable 2 Regulation Line regulation 0 01 3mV Load regulation 0 01 3mV rating current 3A Load regulation 0 02 5mV rating current 3A 3 R...

Page 6: ...for Setting Current Limit INDEPENDENT USE OF MASTER OR SLAVE SUPPLY section of this manual using the MASTER CURRENT control NOTE Because the supplies are being used in series either CURRENT control c...

Page 7: ...R SUPPLY LOAD MASTER SLAVE LOAD _ _ Fig 4 2 Indep endent Operatio n Slave tracking error 0 5 10mV of the master No load with load add Load regulation 300mV B Single supply Fig 4 3 2 6 Meter A Display...

Page 8: ...ORS V 1 C V LABORATORY DC POWER SUPPLY CURRENT POWER 22 13 ON 29 8 7 28 OUTPUT ON O FF GND OFF 12 21 19 20 C V C C PAR SLAVE VOLTAGE GND 17 18 15 14 16 INDEP SERIES PARALLEL TRACKING C C MASTER 3 4 A...

Page 9: ...ator to light 4 Adjust the CURRENT control for the desired current limit Read the current value on the A meter 5 The current limit overload protection has now been preset Do not change the CURRENT con...

Page 10: ...justment of the output voltage of the MASTER supply Also functions as adjustment control for the maximum output voltage of the SLAVE supply when either parallel or series tracking operation Voltage co...

Page 11: ...rity output terminal for the MASTER supply 19 output terminal positive polarity output terminal for the SLAVE supply 17 GND terminal Earth and chassis ground 20 GND terminal Earth and chassis ground 1...

Page 12: ...Fig 3 2 Rear Panel...

Reviews: