              When the series tracking mode of operation is selected, the positive (red) terminal of 

the SLAVE supply output is internally connected to the negative (black) terminal of the 

MASTER supply. 

       In the series tracking mode, the maximum output voltage of both MASTER and 

SLAVE supplies can be simultaneously varied with one control. The maximum SLAVE 

supply voltage is automatically set to the same value as the MASTER supply by using the 

MASTER VOLTAGE controls. 

    A. Set the power supplies to the TRACKING SERIES mode by engaging the left 

TRACKING switch and release the right   

                  TRACKING switch. In this case, the output voltage (across the two supplies) is 

actually double the displayed value.   

         For example, if the MASTER display is set for voltage metering and the SLAVE 

display for current metering, the output voltage across the MASTER positive (red) terminal 

and the SLAVE negative (black) terminal would be double the reading on the MASTER 

LED Display (since both supplies are putting out the same voltage). The actual output 

current would be the value read from the SLAVE LED Display (since the two supplies are 

wired in series, current flowing through each supply must be equal). 

    B . Set the SLAVE CURRENT control the fully clockwise position. The maximum 

current is set using the MASTER CURRENT control. Follow the instructions for "Setting 

Current Limit" (INDEPENDENT USE OF "MASTER" OR "SLAVE" SUPPLY section of 

this manual, using the MASTER CURRENT control. 

NOTE: 

                Because the supplies are being used in series, either CURRENT control can be used 

to set maximum current. If desired, the MASTER CURRENT control can be rotated fully 

clockwise and the SLAVE CURRENT control can be used to adjust the maximum current 

value. Because current through the two supplies must be equal when they are being used in 

series, the lower   

    CURRENT control setting will set the maximum output current. 

        C. Adjust the output voltage the desired level using the MASTER VOLTAGE controls. 

        D. Turn off the power supply and the equipment to be powered during hook-up. 

        E. If "single supply" operation is desired, this allows the power supply to be used as 

twice the voltage and rating current simply by using the negative (black) terminal of    the 

SLAVE supply and the positive (red) terminal of    the MASTER supply, the configuration as 

shown in Fig.4-3. 

               

 

4-4 Operation Mode 

   (1) Independent Operation 

                  The "MASTER" and "SLAVE" supplies each provide a 0 to rating volts output at 

up to rating amps. This procedure covers the use of the MASTER and SLAVE supplies only 

when they are used independently from one another. When used in the INDEPENDENT 

operating mode, the operating controls of the two power supplies are completely 

independent and either supply can be used individually or both can be used simultaneously. 

     A. Disengage both TRACKING mode switches (both switches out) so that the power 

Summary of Contents for Precision AC- DC Power Supply

Page 1: ...Dual Channel 0 30V 5A max or 0 60V single channel AA Portable Power Corp http www batteryspace com Address 860 S 19th St Unit A Richmond CA 94804 Tel 510 525 2328 Fax 510 439 2808 Email Sales battery...

Page 2: ...output range The front panel current controls can be used to establish the output current limit overload or shout circuit when the supply is used as a constant voltage source independent or tracking m...

Page 3: ...utput on off key to enter voltage and current setting C Adjust the output voltage to the desired level using the MASTER VOLTAGE controls D Turn off the power supply and the equipment to be powered dur...

Page 4: ...h its Kindred has not This function is that when the output has been cut off the both intending voltage and current can be preset though in model A B condition only the intending voltage can be adjust...

Page 5: ...peration 1 Output current range 0 rating voltage continuously adjustable 2 Regulation Line regulation 0 01 3mV Load regulation 0 01 3mV rating current 3A Load regulation 0 02 5mV rating current 3A 3 R...

Page 6: ...for Setting Current Limit INDEPENDENT USE OF MASTER OR SLAVE SUPPLY section of this manual using the MASTER CURRENT control NOTE Because the supplies are being used in series either CURRENT control c...

Page 7: ...R SUPPLY LOAD MASTER SLAVE LOAD _ _ Fig 4 2 Indep endent Operatio n Slave tracking error 0 5 10mV of the master No load with load add Load regulation 300mV B Single supply Fig 4 3 2 6 Meter A Display...

Page 8: ...ORS V 1 C V LABORATORY DC POWER SUPPLY CURRENT POWER 22 13 ON 29 8 7 28 OUTPUT ON O FF GND OFF 12 21 19 20 C V C C PAR SLAVE VOLTAGE GND 17 18 15 14 16 INDEP SERIES PARALLEL TRACKING C C MASTER 3 4 A...

Page 9: ...ator to light 4 Adjust the CURRENT control for the desired current limit Read the current value on the A meter 5 The current limit overload protection has now been preset Do not change the CURRENT con...

Page 10: ...justment of the output voltage of the MASTER supply Also functions as adjustment control for the maximum output voltage of the SLAVE supply when either parallel or series tracking operation Voltage co...

Page 11: ...rity output terminal for the MASTER supply 19 output terminal positive polarity output terminal for the SLAVE supply 17 GND terminal Earth and chassis ground 20 GND terminal Earth and chassis ground 1...

Page 12: ...Fig 3 2 Rear Panel...

Reviews: