TRACKING switches. 

      B. Because both voltage and current of the SLAVE supply track the MASTER supply, 

the maximum current and voltage are set using the MASTER controls. Using the MASTER 

supply output jacks, follow the instructions for "Setting Current Limit" (4-2 Section). 

Remember that the actual current output at the MATER supply output jack is double the 

reading on the SLAVE indicator meter. Then push the output on/off key to enter voltage and 

current setting. 

    C. Adjust the output voltage to the desired level using the MASTER VOLTAGE controls. 

    D. Turn off the power supply and the equipment to be powered during hook-up. 

    E. Connect the positive polarity of the device being powered to the red (+) terminal of the 

MASTER power supply. 

    F. Connect the negative polarity of the device being powered to the black (-) terminal of 

the MASTER power supply.   

      The configuration as shown in Fig.4-5. 

 

_

POWER

SUPPLY

M AST ER

S LAVE

+

+

_

Fig .4-5 Paral lel Tracki ng Op eratio n.

L OAD

L OAD

 

 

 

_

POWER

SUPPLY

MAST ER

SL AV E

+

+

_

L OAD

Fig .4-3 Sin gle Supp ly

 

 

Summary of Contents for Precision AC- DC Power Supply

Page 1: ...Dual Channel 0 30V 5A max or 0 60V single channel AA Portable Power Corp http www batteryspace com Address 860 S 19th St Unit A Richmond CA 94804 Tel 510 525 2328 Fax 510 439 2808 Email Sales battery...

Page 2: ...output range The front panel current controls can be used to establish the output current limit overload or shout circuit when the supply is used as a constant voltage source independent or tracking m...

Page 3: ...utput on off key to enter voltage and current setting C Adjust the output voltage to the desired level using the MASTER VOLTAGE controls D Turn off the power supply and the equipment to be powered dur...

Page 4: ...h its Kindred has not This function is that when the output has been cut off the both intending voltage and current can be preset though in model A B condition only the intending voltage can be adjust...

Page 5: ...peration 1 Output current range 0 rating voltage continuously adjustable 2 Regulation Line regulation 0 01 3mV Load regulation 0 01 3mV rating current 3A Load regulation 0 02 5mV rating current 3A 3 R...

Page 6: ...for Setting Current Limit INDEPENDENT USE OF MASTER OR SLAVE SUPPLY section of this manual using the MASTER CURRENT control NOTE Because the supplies are being used in series either CURRENT control c...

Page 7: ...R SUPPLY LOAD MASTER SLAVE LOAD _ _ Fig 4 2 Indep endent Operatio n Slave tracking error 0 5 10mV of the master No load with load add Load regulation 300mV B Single supply Fig 4 3 2 6 Meter A Display...

Page 8: ...ORS V 1 C V LABORATORY DC POWER SUPPLY CURRENT POWER 22 13 ON 29 8 7 28 OUTPUT ON O FF GND OFF 12 21 19 20 C V C C PAR SLAVE VOLTAGE GND 17 18 15 14 16 INDEP SERIES PARALLEL TRACKING C C MASTER 3 4 A...

Page 9: ...ator to light 4 Adjust the CURRENT control for the desired current limit Read the current value on the A meter 5 The current limit overload protection has now been preset Do not change the CURRENT con...

Page 10: ...justment of the output voltage of the MASTER supply Also functions as adjustment control for the maximum output voltage of the SLAVE supply when either parallel or series tracking operation Voltage co...

Page 11: ...rity output terminal for the MASTER supply 19 output terminal positive polarity output terminal for the SLAVE supply 17 GND terminal Earth and chassis ground 20 GND terminal Earth and chassis ground 1...

Page 12: ...Fig 3 2 Rear Panel...

Reviews: