mode, a regulated output voltage is provided. The output voltage remains constant as the load 

increases up until the point where the preset current limit is reached. At that point, the output 

current becomes constant and the output voltage drops in proportion to further increases in load. 

The point is indicated by the front panel LED indicators. The crossover point is reached when 

the CV indicator goes off and the CC indicator comes on. 

              Similarly, crossover from the constant current to the constant voltage mode automatically 

occurs from a decrease in load. A good example of this would be seen when charging a 12-volt 

battery. Initially, the open circuit voltage of the power supply may be preset for 13.8 volts. A low 

battery will place a heavy load on the supply and it will operate in the constant current mode, 

which may be adjusted for a 1 amp charging rate. As the battery becomes charged, and its 

voltage approaches 13.8 volts, its load decreases to the point where it no longer demands the full 

1 amp charging rate. This is the crossover point where the power supply goes into the constant 

voltage mode. 

 

3-1 Front Panel 

22    Power switch: ON/OFF the power input. 

  V LED display: Indicates the MASTER output voltage. 

     

  V LED display: Indicates the SLAVE output voltage. 

     

  A LED display: Indicates the MASTER output current. 

     

  A LED display: Indicates the SLAVE output current. 

     

  Voltage control: for adjustment of the output voltage of the MASTER supply. Also functions 

as adjustment control for the maximum output voltage of the SLAVE supply when either 

parallel or series tracking operation.   

     

 Voltage control: for adjustment of the output voltage of the SLAVE supply when the 

independent operation. 

     

  Current control: for adjustment of the output current of the MASTER supply. Also functions 

as adjustment control for the maximum output voltage of the SLAVE supply when either 

parallel or series tracking operation.   

     

  Current control: for adjustment of the output current of the SLAVE supply.       

     

  C.V indicator: lights when the MASTER supply is in the constant voltage operation, in either 

the Series or Parallel Tracking mode, both the MASTER AND SLAVE supplies are in the 

constant voltage operation. 

      12 C. V indicator: lights when the SLAVE supply is in the constant voltage operation. 

      11 C. C. indicator: lights when the MASTER supply is in the constant current operation. 

      13 C. C. indicator: lights when the SLAVE supply is in the constant current operation. 

                                                            Also lights when the TRACKING PARALLEL mode is selected. 

     

  Over load indicator: lights when load on 5 volt supply becomes too large. 

      14    15    TRACKING Mode Switches: 

      Two push-button switches that select INDEPENDENT mode, series tracking mode, or parallel 

tracking mode as follows: 

      a. When both switches are disengaged (out), the unit is in the INDEPENDENT mode and the 

MASTER and SLAVE power supplies are completely independent from one another. 

            b. When the left switch is engaged (in) and the right switch is disengaged (out), the unit is in 

the TRACKING SERIES mode.   

Summary of Contents for Precision AC- DC Power Supply

Page 1: ...Dual Channel 0 30V 5A max or 0 60V single channel AA Portable Power Corp http www batteryspace com Address 860 S 19th St Unit A Richmond CA 94804 Tel 510 525 2328 Fax 510 439 2808 Email Sales battery...

Page 2: ...output range The front panel current controls can be used to establish the output current limit overload or shout circuit when the supply is used as a constant voltage source independent or tracking m...

Page 3: ...utput on off key to enter voltage and current setting C Adjust the output voltage to the desired level using the MASTER VOLTAGE controls D Turn off the power supply and the equipment to be powered dur...

Page 4: ...h its Kindred has not This function is that when the output has been cut off the both intending voltage and current can be preset though in model A B condition only the intending voltage can be adjust...

Page 5: ...peration 1 Output current range 0 rating voltage continuously adjustable 2 Regulation Line regulation 0 01 3mV Load regulation 0 01 3mV rating current 3A Load regulation 0 02 5mV rating current 3A 3 R...

Page 6: ...for Setting Current Limit INDEPENDENT USE OF MASTER OR SLAVE SUPPLY section of this manual using the MASTER CURRENT control NOTE Because the supplies are being used in series either CURRENT control c...

Page 7: ...R SUPPLY LOAD MASTER SLAVE LOAD _ _ Fig 4 2 Indep endent Operatio n Slave tracking error 0 5 10mV of the master No load with load add Load regulation 300mV B Single supply Fig 4 3 2 6 Meter A Display...

Page 8: ...ORS V 1 C V LABORATORY DC POWER SUPPLY CURRENT POWER 22 13 ON 29 8 7 28 OUTPUT ON O FF GND OFF 12 21 19 20 C V C C PAR SLAVE VOLTAGE GND 17 18 15 14 16 INDEP SERIES PARALLEL TRACKING C C MASTER 3 4 A...

Page 9: ...ator to light 4 Adjust the CURRENT control for the desired current limit Read the current value on the A meter 5 The current limit overload protection has now been preset Do not change the CURRENT con...

Page 10: ...justment of the output voltage of the MASTER supply Also functions as adjustment control for the maximum output voltage of the SLAVE supply when either parallel or series tracking operation Voltage co...

Page 11: ...rity output terminal for the MASTER supply 19 output terminal positive polarity output terminal for the SLAVE supply 17 GND terminal Earth and chassis ground 20 GND terminal Earth and chassis ground 1...

Page 12: ...Fig 3 2 Rear Panel...

Reviews: